Print version of this report
TGA Action ID RC-2025-RN-00845-1
Type of Product Medical Device
Product Name/Description Philips Allura Xper Systems

Allura Xper FD10
Model number: 722003, 722010, 722026

Allura Xper FD10/10
Model number: 722027

Allura Xper FD20
Model number: 722006, 722012, 722028,

Allura Xper FD20/15
Model number: 722058

Allura Xper FD20 OR Table
Model: 722023, 722035

Allura Xper FD20 Biplane
Model number: 722013, 722008
ARTG Number 225815
Unique Device Identifier (UDI) N/A
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action The BIOS battery may deplete faster than initially anticipated. When the BIOS batteryis depleted, the start-up process of the system will be halted and the system will not start. No user messages are shown before the battery is low on power or depleted.

If the problem occurs, the interventional X-ray system will not be available for use. Potential health consequences of the unavailability of the interventional X-ray system relate to a delay of therapy.

Between January 2023 and May 2025, Philips has received 93 complaints potentially related to this problem. Philips has not received any complaints associated with this problem reporting harm.

Philips will replace the BIOS battery in all affected systems.
Instructions for Customers/Users In the case of system failure, customers are to:
- Perform interventions during clinical use that reduce both the likelihood and severity of harm, tailored to relevant hospital or clinical procedures.
- Consider using alternative mobile imaging systems, if available and clinically appropriate.
- Evaluate the feasibility of patient transfer or alternative surgical/ medical treatments, based on the judgement of the healthcare professionals managing the case.

A local Philips representative will arrange for a Field Service Engineer to replace the BIOS battery in the affected systems. This is expected to commence in November 2025.
Action Commencement Date 8/10/2025
Responsible Entity Philips Electronics Australia Ltd
Contact Information 1800 251 400 - Philips Service Delivery Team