TGA Action ID RC-2025-RN-00826-1
Type of Product Medical Device
Product Name/Description Extension Lines for Single Use

Product numbers: DI-100, DI-150, L-303/10

All lot numbers
ARTG Number 443661
Action Type Product Correction
Action Level Hospital
Hazard Classification Class III
Reason for the Action The manufacturer found that if the connection parts come into contact with ethanol-based disinfectants during the usage process, there may be a risk of cracking or leakage of the connection components.

In instances where leakage occurs, there may be a delay in procedure, treatment interruptions, exposure to hazardous drugs, and /or underdosing of medication.

In Australia, BD has received zero reports of this problem, and no adverse event occurred.
Instructions for Customers/Users Customers are to note that this product is sterile; no disinfection is needed.

If other devices connecting to this product require disinfection with ethanol disinfectant, please ensure the product is dry before use.

The manufacturer will update the labelling of all Extension lines for single use to have the following warning added: "This product is sterile; no disinfection is needed. If other devices connecting to this product need disinfection with ethanol disinfectant, ensure it is dry before use."

This action has been closed out on 01/04/2026
Action Commencement Date 12/11/2025
Responsible Entity Becton Dickinson Pty Ltd
Contact Information 1800 656 100 - Customer Service Team