| TGA Action ID |
RC-2025-RN-00824-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
IV Gravity Burette Set
Description: 53 cm (21") Add-On 150 ml Burette Set (Clave, Shut-Off), Vented Cap
Product Code: 011-C7014
Lot Numbers: 14029638, 14065820, 14110075, 14129823 |
| ARTG Number |
269616 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
ICU Medical has identified a lot-specific problem where the Burette component of the IV Gravity Burette Set is missing an internal shutoff valve intended to stop fluid flow. The affected products will be missing the blue shutoff valve within the burette chamber.
If a device without the shutoff valve is used, it may result in over delivery/unrestricted flow or air may be infused into the body. This may also result in a delay in therapy during setup. To date, ICU Medical has not received reports of serious injury or deaths associated with this problem. |
| Instructions for Customers/Users |
Customers are to: - Check all inventory and quarantine affected lots. - Destroy all affected products. - ICU Medical will provide replacement products upon receipt of Customer Response Form (sent to customers). |
| Action Commencement Date |
30/09/2025 |
| Responsible Entity |
|
| Contact Information |
1300 428 652 - ICU Medical Customer Service |