Print version of this report
TGA Action ID RC-2025-RN-00821-1
Type of Product Medical Device
Product Name/Description IMMULITE 2000 Intact PTH. An in vitro diagnostic medical device (IVD).

Test Code: iPT

Siemens Material Number/Catalog Number: 10381441, 10381442, L2KPP2, L2KPP6, 0381441, 10381442, L2KPP2,

Unique Device Identification: 00630414961910, 00630414961927

Lot Numbers: 385, 386

Expiry Dates: 17/12/2025, 05/03/2026
ARTG Number 179720
Unique Device Identifier (UDI) 00630414961910, 00630414961927
Action Type Recall
Action Level Hospital
Hazard Classification Class III
Reason for the Action Siemens Healthineers has identified the potential for falsely depressed Intact PTH patient results at the low end of the assay range, = 50 pg/mL (= 5.3 pmol/L), when using the affected lots of IMMULITE 2000/IMMULITE 2000 XPi systems.

The bias is observed in both serum and plasma samples.

This defect could potentially result in additional testing being required.

There are no unaffected kit lots available. Siemens is working to restore assay performance and a follow-up communication will be provided once restored.
Instructions for Customers/Users Customers are to:
- Review the customer letter with a Medical Director to determine the appropriate course of action, including previously generated results.
- Discontinue use of and discard the affected lots.
- Contact local Siemens Healthineers Account Manager, Customer Care Centre, or Technical Support Representative to discuss suitable alternatives.
Action Commencement Date 1/10/2025
Responsible Entity Siemens Healthcare Pty Ltd
Contact Information 1800 310 300 - Siemens Customer Care Centre