| TGA Action ID |
RC-2025-RN-00821-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
IMMULITE 2000 Intact PTH. An in vitro diagnostic medical device (IVD).
Test Code: iPT
Siemens Material Number/Catalog Number: 10381441, 10381442, L2KPP2, L2KPP6, 0381441, 10381442, L2KPP2,
Unique Device Identification: 00630414961910, 00630414961927
Lot Numbers: 385, 386
Expiry Dates: 17/12/2025, 05/03/2026 |
| ARTG Number |
179720 |
| Unique Device Identifier (UDI) |
00630414961910, 00630414961927 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class III |
| Reason for the Action |
Siemens Healthineers has identified the potential for falsely depressed Intact PTH patient results at the low end of the assay range, = 50 pg/mL (= 5.3 pmol/L), when using the affected lots of IMMULITE 2000/IMMULITE 2000 XPi systems.
The bias is observed in both serum and plasma samples.
This defect could potentially result in additional testing being required.
There are no unaffected kit lots available. Siemens is working to restore assay performance and a follow-up communication will be provided once restored. |
| Instructions for Customers/Users |
Customers are to: - Review the customer letter with a Medical Director to determine the appropriate course of action, including previously generated results. - Discontinue use of and discard the affected lots. - Contact local Siemens Healthineers Account Manager, Customer Care Centre, or Technical Support Representative to discuss suitable alternatives. |
| Action Commencement Date |
1/10/2025 |
| Responsible Entity |
|
| Contact Information |
1800 310 300 - Siemens Customer Care Centre |