TGA Action ID RC-2025-RN-00819-1
Type of Product Medical Device
Product Name/Description Leksell Vantage Arc System and Leksell Vantage Support Arc

Serial numbers: SH00001-SH00338
ARTG Number 509585
Unique Device Identifier (UDI) Arc: 07340048304856, Support Arc: 07340048304849
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Elekta is notifying that for specified serial numbers, (SH00001-SH00338) there is an increased risk that paint flakes from the Leksell Vantage Arc system ‘s engravings may separate from the Leksell Vantage Arc System components. Separation of paint flakes is likely to occur during cleaning and sterilisation.

If paint flakes are not washed away during the cleaning and sterilisation process, these could remain on the Leksell Vantage Arc System when used in surgery. This poses a risk of paint flakes entering the surgical area.
This situation has not been observed in clinical practice, but the risk should be fully considered and mitigated. The paint does not include any metallic components and is not magnetic.

In a worst-case scenario, an undetected sterile fragment of non-metallic paint may inadvertently enter the wound. Potential risks associated with this occurrence include infection, inflammation, delayed healing, and possible migration into surrounding tissues which may result in further complications depending on the affected area.

The likelihood of this hazard occurring is considered low, as there have been no reports of it being observed in clinical practice.
Instructions for Customers/Users To prevent the risk of paint flakes entering the surgical area, it is recommended to carefully inspect all the components of the Arc system for any loose paint flakes after cleaning and before use. If any loose paint flakes are found, thoroughly wipe the Arc components with a soft, lint-free cloth dampened with alcohol or distilled water until no loose paint flakes are present on the product. After wiping, ensure that no paint flakes remain on the gloves of the person who performed the cleaning.
If customers experience any further issues with paint flaking, contact Elekta for assistance. The representative will assess the situation and determine if a replacement device is necessary.

This action has been finalised on 6/3/2026.