| TGA Action ID |
RC-2025-RN-00806-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Sharesource Connectivity Platform for use with Homechoice Claria APD System - Version 8.13.6
Product code: 5CGM01 |
| ARTG Number |
266804 |
| Unique Device Identifier (UDI) |
15413765545723 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Vantive has identified a software defect within the Sharesource Claria application. If a clinical user attempts to update a patient’s program (prescription settings) while using multiple internet browser windows simultaneously, the changes do not update on that patient’s Homechoice Claria APD cycler located with the patient.
Performing therapy using an outdated prescription may result in a patient receiving more therapy or less therapy than intended. There have been no reports of adverse events or serious injuries related to this problem. |
| Instructions for Customers/Users |
Vantive will correct this problem with a software update.
Vantive will contact affected customers directly to provide specific instructions on how to correct those specific patient programs.
This action has been closed out on 22/05/2026 |
| Action Commencement Date |
24/09/2025 |
| Responsible Entity |
|
| Contact Information |
1800 920 133 - Customer Service |