TGA Action ID RC-2025-RN-00797-1
Type of Product Medical Device
Product Name/Description Hugo RAS system - Initial Launch Instruments

Model Numbers: MRASI0007, MRASI0012
GTIN: 10884521740006, 10884521740051

Year (First 3 Digits): C21
ID (Second 3 Digits): BAE, BAH
Range (Last 4 Digits): All

All serial numbers
ARTG Number 377600, 378145
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action When an Initial Launch Instrument is attached to a Hugo RAS system running software version 2.2 or higher, the system may display the message: “Instrument expired. Detach instrument and replace.” This problem occurs upon attachment and can be immediately resolved by replacing the affected instrument.

Medtronic has received 9 complaints related to this problem, with no reports of harm to either the patient or user. The potential risks include the inability to use an affected instrument, which could lead to a delay in treatment (i.e. prolonged procedure) and/or bleeding.
Instructions for Customers/Users Customers are to:

• immediately quarantine and discontinue use of the affected Initial Launch Instruments with associated serial numbers, listed in Appendix A of the customer letter.
• return all affected products. Medtronic will assist with the return of affected products upon request.

Medtronic will work with customers to ensure that replacement instruments are available to prevent interruptions in patient care.

This action has been finalised on 21/01/2026.
Action Commencement Date 16/09/2025
Responsible Entity Medtronic Australasia Pty Ltd
Contact Information 0486 006 068 - Glen Kuhn, Medtronic Australasia