Print version of this report
TGA Action ID RC-2025-RN-00796-1
Type of Product Medical Device
Product Name/Description CONSTELLATION ULTRAVIT (10K) and HYPERVIT (20K) Probes

Multiple material and product codes
ARTG Number 145666, 319177, 370933
Action Type Product Alert
Action Level Hospital
Hazard Classification Class II
Reason for the Action There is potential for specific models of ULTRAVIT and HYPERVIT to unexpectedly fail to actuate and cut during use. These probes are intended for use with the CONSTELLATION Vision System and packs or kits.
Instructions for Customers/Users To mitigate the potential risk associated with unexpected cut stoppage during surgery, Alcon is advising customers to take the following precautions:

1. Strictly adhere to the IFU regarding limitations, including maximum use time and single use only.
2. When using a probe model identified as within the scope:
a. Reduce the actuation rate to no more than 5000 actuations per minute.
b. Adjust other related console settings as needed to accommodate the reduced cutting/actuation rate.
c. If a reduction of cutting capability or failure to actuate is observed during the surgical procedure, stop immediately. Take all necessary surgical precautions to remove the probe and replace it with an unused probe.
d. Report any faulty probe identified.

This action has been closed out on 18/12/2025
Action Commencement Date 17/09/2025
Responsible Entity Alcon Laboratories Australia Pty Ltd
Contact Information 1800 224 153 - Alcon’s Surgical Customer Services