Print version of this report
TGA Action ID RC-2025-RN-00782-1
Type of Product Medical Device
Product Name/Description Fisher & Paykel Healthcare Airvo 2 and myAirvo 2 Humidifier

Product Code: PT100AN, PT101AN

Device Identifier: 09420012447319, 09420012447333
ARTG Number 177298
Action Type Product Correction
Action Level Consumer
Hazard Classification Class II
Reason for the Action The Airvo 2 and myAirvo 2 device has an audible alert that sounds for at least 120 seconds when the unit is unintentionally disconnected from a power source. The Disinfection Kit User Manual has been updated to include a regular test of this alert, to be carried out in between patient uses.

Fisher & Paykel (F&P) have updated the manual due to reports of the audible alert sounding for less than 120 seconds, which may delay user awareness to reconnect power.
Instructions for Customers/Users F&P have updated the Disinfection Kit User Manual with a step to check the audible alert.

Users may continue to use Airvo 2 and myAirvo 2 devices with appropriate patient monitoring as per the User Manual. When using the device, all instructions, including warnings and cautions in the User Manual must be followed.

This action has been finalised on 16/12/2025.
Action Commencement Date 15/09/2025
Responsible Entity Fisher & Paykel Healthcare Pty Ltd
Contact Information 03 9871 4900 - Fisher & Paykel Healthcare Australia Customer Care