| TGA Action ID |
RC-2025-RN-00781-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Azurion R2.x or R3.x systems
Model numbers: 722078, 722079, 722223, 722224, 722226, 722228, 722233, 722236 |
| ARTG Number |
225815, 370112 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class III |
| Reason for the Action |
Potential temporary loss of motorised movements due to firewall router not installed.
If this router is not installed, there could be interference between the Azurion and the hospital network. When this occurs, the Azurion will initiate an automatic restart of the positioning software, during which all motorised movements will be disabled and the message “The geometry is restarting” will be displayed on the screen. The restart will take approximately 90 seconds to complete, after which motorised movements will be fully available.
X-ray imaging is not impacted by this problem and remains available during the positioning software restart.
Philips has received no reports of patient harm related to this problem. |
| Instructions for Customers/Users |
Customers should circulate this Product Correction to all users so that they are aware of the problem.
Philips will inspect all specific Azurion systems to confirm that the firewall router is installed and if not, Philips will be installing the firewall router free of charge. Philips expects to start this activity in December 2025. Your local Philips representative will contact you to schedule this activity.
This action has been closed out on 12/11/2026 |
| Action Commencement Date |
10/09/2025 |
| Responsible Entity |
|
| Contact Information |
1800 251 400 - Philips Service Delivery Team |