TGA Action ID RC-2025-RN-00780-1
Type of Product Medical Device
Product Name/Description THA 5.0, TKA 2.1, PKA 3.1 Install Disk deployed on Mako 4.0

Part Number 229999

Catalogue Number: 700003123526

Lot Numbers: 0003526, 0004526, 0005526

GTIN: 07613327650990
ARTG Number 223106
Action Type Product Alert
Action Level Hospital
Hazard Classification Class II
Reason for the Action Stryker discovered that a past research and development implant, Restoris Porous (MAKO Surgical Corp.), is an available option in the implant selection dropdown for the Partial Knee Arthroplasty (PKA) 3.1 application. The software displays a non-supported implant system within the PKA 3.1 application.

The default implant system selection in the PKA 3.1 application is Restoris MCK. If the user switches from the default implant system selection to the unsupported implant system Restoris Porous, they will be required to revert to MCK and complete the MCK bone prep as planned. This may result in a surgical delay of up to 15 minutes.
Instructions for Customers/Users Customers are to following the mitigations that already exist within the software application:

·The default implant selection in the PKA 3.1 application is Restoris MCK. The user/surgeon would have to manually change the selection to Restoris Porous implant system on the PKA user interface and acknowledge/agree to the implant system selection prior to surgery.
·The surgeon’s acknowledgement indicates they have read the PKA 3.1 User Guide and Surgical Techniques for the selected implant system. The Restoris Porous implant system is not listed in the User Guide and does not have a Surgical Technique available. This should alert the surgeon that they should not proceed with the Restoris Porous implant system selection.
·The cut sequences during bone preparation will also act as a safeguard to detect the incorrect implant system. The cut sequence for Restoris Porous requires a specific instrument called a Tier Drill tool to be used. The Tier Drill, trials and implants associated with Restoris Porous are not offered by Stryker, therefore the surgeon will be unable to proceed with the Restoris Porous implant system selected.

Customers will be notified once a software correction is available.

This action has been closed out on 06/08/2026
Action Commencement Date 10/09/2025
Responsible Entity Stryker Australia Pty Ltd
Contact Information postmarketssp@stryker.com - Stryker Australia