| TGA Action ID |
RC-2025-RN-00778-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
TactiFlex Ablation Catheter, Sensor Enabled
TactiFlex Ablation Catheter with Uni-Directional Handle, Sensor Enabled Models: A-TFSE-D, A-TFSE-F, A-TFSE-J
TactiFlex Ablation Catheter with Bi-Directional Handle, Sensor Enabled Models: A-TFSE-DD, A-TFSE-DF, A-TFSE-FF, A-TFSE-FJ, A-TFSE-JJ |
| ARTG Number |
380981, 380982 |
| Unique Device Identifier (UDI) |
TactiFlex Ablation Catheter with Uni-Directional Handle, Sensor Enabled, TactiFlex Ablation Catheter with Bi-Directional Handle, Sensor Enabled |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Abbott has become aware of a limited number of tip detachment events during catheter removal from the packaging. Although the packaging has not changed since the introduction of the device (2022), improper removal from the packaging can result in tip damage.
Abbott has experienced 3 recent instances where tips have detached and remained in the patient. In all three (3) cases there were no further adverse outcomes reported to the patients. |
| Instructions for Customers/Users |
Customers should: * Continue following the product IFU by inspecting the packaging and catheter for damage and/or kinking prior to use. * Review the guidance provided in the customer letter in Appendix A information on proper removal technique of the TactiFlex Ablation Catheter, Sensor Enabled from the packaging tray. * Be aware that Abbott has made changes to the packaging tray design to help mitigate tip damage during catheter removal from the packaging. The deployment of the new packaging has begun.
This action has been closed out on 28/10/2025 |
| Action Commencement Date |
11/09/2025 |
| Responsible Entity |
|
| Contact Information |
0488 264 626 - Joanne Newling, Abbott Medical Australia |