TGA Action ID RC-2025-RN-00774-1
Type of Product Medical Device
Product Name/Description CellMek SPS Sample Preparation System. An in vitro diagnostic medical device (IVD)

Reference number: C44603

All manufactured units
ARTG Number 177999
Unique Device Identifier (UDI) 15099590750312
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Two problems have been identified that may cause incorrect sample preparation and result in incorrect results from the flow cytometric analysis.

Problem 1:
The dead volumes in Appendix D, Table D.1, in the CellMek SPS Instructions for Use (IFU), for Cassette Types B and E are incorrect. These cassette types are used only with Sarstedt brand tubes.
• Cassette Type B: IFU states 750µl, but actual dead volume is 1582µl; Used with 11mm and 13mm diameter Sarstedt brand tubes.
• Cassette Type E: IFU states 1300µl, but actual dead volume is 2093µl; Used with 15mm diameter Sarstedt brand tubes.

Problem 2:
There's an anomaly with CellMek SPS software that causes the system not to rock Cassette Type E before tubes are pierced for sample preparation. Cassette Type E is used with 15mm X 92mm Sarstedt brand tubes. Rocking the cassettes prior to piercing is needed to ensure specimen homogeneity.
Instructions for Customers/Users Beckman Coulter will be amending the IFU in the next revision and will address the software anomaly in a future product release.

Meanwhile, customers are read the Customer Letter and:
- For problem 1, use the actual specimen dead volumes stated in Problem Description in the customer letter when processing samples in CellMek SPS with Sarstedt brand tubes.
- For problem 2, not use Cassette Type E (15mm X 92mm Sarstedt brand tubes) to process samples on CellMek SPS.

This action has been closed out on 28/10/2026
Action Commencement Date 11/09/2025
Responsible Entity Beckman Coulter Australia Pty Ltd
Contact Information 1800 060 878 - Beckman Coulter Australia Customer Service