TGA Action ID RC-2025-RN-00773-1
Type of Product Medical Device
Product Name/Description ID-Anti-S.

ID-Anti-S
UDI: 07611969069514
Catalog Number: 007132
Batch/Lot: 0033454701
Expiry date: 31.07.2026

ID-Anti-M, N, S, s, Fya, Fyb
UDI: 07611969069552
Catalog Number: 008712
Batch/Lot: 0050103803, 0076423804, 0076443805, 0076463806
Expiry date: 30.04.2026

An in vitro diagnostic medical device (IVD)
ARTG Number 217696
Unique Device Identifier (UDI) 07611969069514, 07611969069552
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Bio-Rad has received complaints of false positive reactions associated with ID-Anti-S. An internal investigation confirmed that the test serum ID-Anti-S was contaminated with an anti-Dia antibody. The contaminant was identified in a raw material used during the manufacture of the affected lots. The presence of the contaminant may lead to false positive results.

In cases where a false positive result occurs, treatment may be delayed. However, the associated clinical impact is generally considered minor, typically resulting in temporary discomfort or transient conditions. Such instances usually necessitate the collection of an additional sample to resolve the discrepancy.
Instructions for Customers/Users Customers are to:
- Check inventory for any remaining affected lots and dispose following internal protocol.
- Contact Bio-Rad to request replacement for any of the affected products.
- If any patients or donors were tested with the affected lots and received a positive result, consult a medical director to decide if further testing or other steps are needed.

This action has been finalised on 22/12/2025.
Action Commencement Date 9/09/2025
Responsible Entity Bio-Rad Laboratories Pty Ltd
Contact Information (02) 9914 2815 - Bio-Rad Laboratories