| TGA Action ID |
RC-2025-RN-00773-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
ID-Anti-S.
ID-Anti-S UDI: 07611969069514 Catalog Number: 007132 Batch/Lot: 0033454701 Expiry date: 31.07.2026
ID-Anti-M, N, S, s, Fya, Fyb UDI: 07611969069552 Catalog Number: 008712 Batch/Lot: 0050103803, 0076423804, 0076443805, 0076463806 Expiry date: 30.04.2026
An in vitro diagnostic medical device (IVD) |
| ARTG Number |
217696 |
| Unique Device Identifier (UDI) |
07611969069514, 07611969069552 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Bio-Rad has received complaints of false positive reactions associated with ID-Anti-S. An internal investigation confirmed that the test serum ID-Anti-S was contaminated with an anti-Dia antibody. The contaminant was identified in a raw material used during the manufacture of the affected lots. The presence of the contaminant may lead to false positive results.
In cases where a false positive result occurs, treatment may be delayed. However, the associated clinical impact is generally considered minor, typically resulting in temporary discomfort or transient conditions. Such instances usually necessitate the collection of an additional sample to resolve the discrepancy. |
| Instructions for Customers/Users |
Customers are to: - Check inventory for any remaining affected lots and dispose following internal protocol. - Contact Bio-Rad to request replacement for any of the affected products. - If any patients or donors were tested with the affected lots and received a positive result, consult a medical director to decide if further testing or other steps are needed.
This action has been finalised on 22/12/2025. |
| Action Commencement Date |
9/09/2025 |
| Responsible Entity |
|
| Contact Information |
(02) 9914 2815 - Bio-Rad Laboratories |