| Reason for the Action |
Olympus previously issued a Product Correction and Safety Alert in 2023, after investigating complaints of endobronchial combustion during therapeutic bronchoscopy.
Since the Corrective Action in 2023, Olympus conducted additional assessment on the use of compatible bronchoscopes with laser, argon plasma coagulation, and high-frequency therapy equipment. Through these investigations, Olympus has determined that additional IFU updates are necessary to provide further clarification on safe and effective use of the bronchoscopes when used in conjunction with these instruments. This Correction supersedes the IFU addendum provided in 2023.
If endobronchial combustion occurs, patients may suffer critical internal burns to the airway or lungs that may result in a requirement for additional medical intervention, prolonged procedure, extended hospitalisation or ICU care, and death. Combustion can also result in damage to or breakage of device components that may injure or remain unintendedly in the patient and/or may require retrieval or surgical removal. |