TGA Action ID RC-2025-RN-00768-1
Type of Product Medical Device
Product Name/Description Olympus Bronchofibrescopes and Bronchovideoscopes

Multiple models, catalogue numbers and serial numbers
ARTG Number 112269, 145873
Action Type Product Correction
Action Level Hospital
Hazard Classification Class I
Reason for the Action Olympus previously issued a Product Correction and Safety Alert in 2023, after investigating complaints of endobronchial combustion during therapeutic bronchoscopy.

Since the Corrective Action in 2023, Olympus conducted additional assessment on the use of compatible bronchoscopes with laser, argon plasma coagulation, and high-frequency therapy equipment. Through these investigations, Olympus has determined that additional IFU updates are necessary to provide further clarification on safe and effective use of the bronchoscopes when used in conjunction with these instruments. This Correction supersedes the IFU addendum provided in 2023.

If endobronchial combustion occurs, patients may suffer critical internal burns to the airway or lungs that may result in a requirement for additional medical intervention, prolonged procedure, extended hospitalisation or ICU care, and death. Combustion can also result in damage to or breakage of device components that may injure or remain unintendedly in the patient and/or may require retrieval or surgical removal.
Instructions for Customers/Users Customers are to:
- Inspect inventory for affected devices
- Review updated IFU and discard any previous IFUs
- For products that have been discontinued place a copy of the customer letter with the updated wording with the current IFU

This action has been finalised on 21/01/2026.
Action Commencement Date 16/09/2025
Responsible Entity Olympus Australia Pty Ltd
Contact Information 1300 132 992 - Olympus Customer Operations