| TGA Action ID |
RC-2025-RN-00766-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
VYNTUS products with SNIP/RHINO Option
Optional Kit P/N 708052
SNIP Option for Vyntus PNEUMO/IOS/APS Model: V-570303
SNIP Option for Vyntus ONE/BODY Model: V-578603
Rhinomanometry Option for Vyntus PNEUMO/IOS/ APS Model: *V-570304
Rhinomanometry Option for Vyntus BODY/ONE Model: V-578604
*Optional Kit P/N 706188 incudes Elbow 45° with tube and adapter part for Vyntus Pneumo only |
| ARTG Number |
156954 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
The hygiene manual does not contain proper reprocessing instructions for the cleaning, disinfection, and disposal of some components (V-708052 Connection set+ 706188 Elbow 45° with tubing) associated with these testing features.
Without proper cleaning instructions, clinicians may reuse parts intended for single use. As a result, a contaminated part used on an infected patient can transmit pathogens to the next patient requiring the same test and parts, leading to cross-contamination and potential exposure to infection. |
| Instructions for Customers/Users |
An update will be made to the Instructions for Use (IFU), to ensure there are correct instructions regarding reprocessing and all impacted components are described as single use. The updated IFU will be available starting in October 2025.
This action has been finalised on 22/12/2025. |
| Action Commencement Date |
5/09/2025 |
| Responsible Entity |
|
| Contact Information |
02 9006 1662 - Emergo Australia |