TGA Action ID RC-2025-RN-00766-1
Type of Product Medical Device
Product Name/Description VYNTUS products with SNIP/RHINO Option

Optional Kit P/N 708052

SNIP Option for Vyntus PNEUMO/IOS/APS
Model: V-570303

SNIP Option for Vyntus ONE/BODY
Model: V-578603

Rhinomanometry Option for Vyntus PNEUMO/IOS/ APS
Model: *V-570304

Rhinomanometry Option for Vyntus BODY/ONE
Model: V-578604

*Optional Kit P/N 706188 incudes Elbow 45° with tube and adapter part for Vyntus Pneumo only
ARTG Number 156954
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action The hygiene manual does not contain proper reprocessing instructions for the cleaning, disinfection, and disposal of some components (V-708052 Connection set+ 706188 Elbow 45° with tubing) associated with these testing features.

Without proper cleaning instructions, clinicians may reuse parts intended for single use. As a result, a contaminated part used on an infected patient can transmit pathogens to the next patient requiring the same test and parts, leading to cross-contamination and potential exposure to infection.
Instructions for Customers/Users An update will be made to the Instructions for Use (IFU), to ensure there are correct instructions regarding reprocessing and all impacted components are described as single use. The updated IFU will be available starting in October 2025.

This action has been finalised on 22/12/2025.
Action Commencement Date 5/09/2025
Responsible Entity Emergo Asia Pacific Pty Ltd T/a Emergo Australia