TGA Action ID RC-2025-RN-00761-1
Type of Product Medical Device
Product Name/Description Patient monitor Vista 120/Vista120s
ARTG Number 294160, 314265
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Under certain conditions, the monitor can temporarily freeze without an alarm or restart being triggered.

Analyses have shown that such screen freeze may happen if e.g. the users deviate from the instructions for use. In particular, the freeze may occur if the Vista 120 detects a damaged or full storage device. If so, the monitor will display a corresponding prompt/ icon.
Instructions for Customers/Users Customers should read and affix the supplement, contained within the customer letter, to their instructions for use and inform both all staff engaged in the device of the supplement and of the necessity to comply with the amended instructions.

This action has been finalised on 19/12/2025.
Action Commencement Date 4/09/2025
Responsible Entity Draeger Australia Pty Ltd
Contact Information 1800 372 437 - Customer service