| Product Name/Description |
Philips Azurion and Allura with 1 Phase UPS
Product Names: Allura Xper FD10, Allura Xper FD10/10, Allura Xper FD20 OR Table, Azurion 7 B12, Azurion 7 B20, Azurion 7 M12, Azurion 7 M20, Azurion 3 M12, Azurion 7 M12, Azurion 7 M20, Azurion 7 B20, Azurion 5 M12, Azurion 5 M20
Product Codes: 722026, 722027, 722035, 722067, 722068, 722078, 722079, 722221, 722223, 722224, 722226, 722227, 722228 |
| Reason for the Action |
Philips has identified that some internal components of the 1 Phase UPS may fail. When this occurs, there will be a complete loss of power to the system, causing the system to shut down and/or not start up. The 1 Phase UPS system may fail without warning and prior to any mains power failure.
If the system does not start up or the system shuts down, there is a potential risk of delay or termination of the procedure. The potential delay in treatment and/or termination of the procedure may result in serious adverse health outcomes, including the possibility of death, especially when the system is used with critical patients.
In the period from January 2022 to May 2025, Philips has received 153 complaints related to Azurion systems and 32 complaints related to Allura systems associated with this problem. None of these complaints reported any harm to the patient. |