TGA Action ID RC-2025-RN-00748-1
Type of Product Medical Device
Product Name/Description C-MAC 8404XXX Video Laryngoscope

Multiple Product Descriptions

Material Numbers: 8404KXC, 8404AXC, 8404BXC, 8404HX, 8404EXC, 8404HXP, 8404DXC, 8404GXC, 8404MXC

RI Numbers: AC81AU, AC79AU, AC7BU, AC7FAU, AC7DAU, C80AU, AC7CAU, AC7EAU, AC82AU
ARTG Number 285413
Unique Device Identifier (UDI) 4048551001493U8
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action The C-MAC 8404XXX Video Laryngoscope reprocessing instructions (RI) for Australia have been updated to indicate the number of verified cycles for ‘Automated Cleaning: Thermal Disinfection’ and ‘Sterilisation: Low Temperature’.

The RI’s now mention that 200 cycles of ‘Automated Cleaning: Thermal Disinfection’ have been verified and 45 cycles of ‘Sterilisation: Low Temperature’ have been verified for the C-MAC 8404XXX Video Laryngoscopes. These devices are unable to withstand long term reprocessing with ‘Sterilisation: Low Temperature’.

The use of these reprocessing methods as well as other variables including chemical properties, temperature, frequency and handling can have an impact on the lifespan of the C-MAC Video Laryngoscope. A visual and functional check must be carried out in accordance with the RI and Instructions for Use before and after use.
Instructions for Customers/Users Customers are to:
- Discard any previous revision of the Reprocessing Instruction (RI).
- Use the updated RI attached to the customer letter.

KARL STORZ recommends ‘Automated Cleaning: Thermal Disinfection’ as the preferred final stage of reprocessing for the C-MAC 8404XXX Video Laryngoscopes and the use of ‘Sterilisation: Low Temperature’ as an optional method.
Action Commencement Date 28/08/2025
Responsible Entity Karl Storz Endoscopy Australia Pty Ltd
Contact Information 1800 996 562 - Karl Storz Customer Service Line