| TGA Action ID |
RC-2025-RN-00746-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
CORFLO Nasogastric / Nasointestinal Feeding Tubes and Cortrak 2 Nasogastric / Nasointestinal Feeding Tube with Electromagnetic Transmitting Stylet
Multiple Product Codes
All Serial Numbers |
| ARTG Number |
295243, 297721 |
| Action Type |
Product Correction |
| Action Level |
Consumer |
| Hazard Classification |
Class II |
| Reason for the Action |
Avanos Medical Inc. has received reports from customers experiencing problems of unresolvable clogging, bursting, and/or tearing when using Nasogastric (NG) tubes.
Possible outcomes that could be as a result of use or misuse of any feeding tube or its replacement are: • Pneumothorax • GI perforation • Aspiration • Airway obstruction • Tissue irritation or necrosis • Allergic reaction • Contamination • Delay in diagnosis • Delay or miss dose of medication or nutrition and related complications or the need for additional medical procedures. |
| Instructions for Customers/Users |
Customers are to read and follow the updated IFU which includes guidance for the device lifetime, tube maintenance and unclogging recommendations.
A copy can be downloaded here eIFU - Avanos Medical Inc or call Avanos for a copy.
This action has been finalised on 18/12/2025. |