TGA Action ID RC-2025-RN-00745-1
Type of Product Medical Device
Product Name/Description Space Perfusor and SpacePlus Perfusor

Space Perfusor
Article Code: 34521041
Batches: 92204627, 92204263

SpacePlus Perfusor
Article Code: 34521040, 34523142
Batches: 97088537, 97089719, 97088136
ARTG Number 114011
Action Type Product Correction
Action Level Hospital
Hazard Classification Class I
Reason for the Action During routine in-process inspections, certain components of the drive unit from specific spare part batches showed geometric deviations that exceed the defined manufacturing tolerances.

Under normal conditions, contact between the drive head and plunger plate triggers an infrared sensor that enables controlled deceleration and claw engagement.

In affected units, a geometric deviation prevents the membrane plate from retracting, which causes the infrared sensor to fail and the drive head to move uncontrollably. Depending on the medication administered, this could lead to serious harm, including the risk of fatal outcomes.
Instructions for Customers/Users Customers are to:

- Identify impacted batch numbers according to the instructions included.
- Quarantine affected devices and return affected parts

B Braun Australia will facilitate a free of charge repair or spare part replacement

This action has been closed out on 30/06/2026
Action Commencement Date 27/08/2025
Responsible Entity B Braun Australia Pty Ltd
Contact Information 1800 251 705 - B Braun Technical Service