TGA Action ID RC-2025-RN-00743-1
Type of Product Medical Device
Product Name/Description System 7/8/9 Reciprocating Saw & System 8 Sternum Saw

Catalogue Numbers: 7206000000, 8206000000, 9206000000, 8207000000

GTIN Numbers: 07613327064834, 07613327304473, 07613327142716, 07613327304398

Multiple lot numbers
ARTG Number 285715
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action There is a potential for the Compression Spring (0038-419-000) inside the saw’s drivetrain assembly to disengage during blade loading and unloading. A disengaged spring disables the blade collar from springing back to its original home position, potentially leading to blade retention failure when loading a blade.
Instructions for Customers/Users Customers are to:
1. Immediately inspect inventory to locate and quarantine any affected products they have.
2. Be aware that Stryker’s ProCare team will contact affected customers to arrange for affected devices to be repaired. The devices will be returned to customers upon completion of the repair.


This action has been closed out on 04/03/2026.
Action Commencement Date 26/08/2025
Responsible Entity Stryker Australia Pty Ltd