TGA Action ID RC-2025-RN-00740-1
Type of Product Medical Device
Product Name/Description HeartSine Pad-Pak used with HeartSine SAM 350P, HeartSine SAM 360P, HeartSine SAM 500P

Pad-Pak Catalogue Numbers: PADPAK03, PADPAK04, PADPAK07

HeartSine Catalogue Numbers: 350BASUK10, 360BASSJ10, 360BASUK10, 500BASUK10, 500BASUKGW

Expiry dates: 17 April 2025 to 01 August 2029
ARTG Number 312730, 338960
Action Type Product Correction
Action Level Consumer
Hazard Classification Class I
Reason for the Action Post-market surveillance has revealed that the Pad-Paks are not always inserted properly into the HeartSine AEDs as outlined in the User Manual, which may cause failure during device use. Upon investigation, the potential causes of the improper insertion of Pad-Paks include user error and bent locator pins, which may occur during the manufacturing process.

In the event the device is unable to complete electrical connection via the pins, the device will repeatedly prompt “apply pads to patient’s bare chest”. In some cases, the HeartSine AED may fail to power on entirely.

If the Pad-Pak is not properly inserted into the HeartSine AED, or if the Pad-Pak locator pins are bent, the device may fail to deliver the intended therapy during use, potentially leading to a delay in treatment or no treatment being delivered during use.

Since 2018, there have been 36 adverse events reported globally. Of these, 5 were confirmed to be caused by bent locator pins, 2 devices were not returned for investigation and 29 were determined to be related to potential user error.
Instructions for Customers/Users End users and distributors are to:
- Inspect any Pad-Paks on hand to determine if the expiry is between 17 April 2025 to 01 August 2029.
- If within the expiry date range follow the instructions in the customer letter and remove the Pad-Pak and check for bent locator pins.
- If locator pins are straight, re-insert the Pad-Pak correctly.
- If the locater pins are bent, remove that Pad-Pak and set it aside.
- If the locater pins are bent, Stryker will provide replacement Pad-Paks, once the response form is received.

Further instructions are provided in the customer letter.
Action Commencement Date 17/09/2025
Responsible Entity Stryker Australia Pty Ltd