TGA Action ID RC-2025-RN-00730-1
Type of Product Medical Device
Product Name/Description Exactech Equinoxe Reverse Shoulder Humeral Liners

REVERSE 38MM HUMERAL LINER +0
Cataogue Number: 320-38-00

REVERSE 38MM HUMERAL LINER +2.5
Cataogue Number: 320-38-03

REVERSE 42MM HUMERAL LINER +0
Cataogue Number: 320-42-00

REVERSE 42MM HUMERAL LINER +2.5
Cataogue Number: 320-42-03

Multiple serial numbers
ARTG Number 264172
Unique Device Identifier (UDI) Equinoxe Humeral Liner
Action Type Recall / Product Alert
Action Level Hospital
Hazard Classification Class II
Reason for the Action The manufacturer has identified that impacted reverse shoulder humeral liners may have an articular surface positioned up to 0.5 mm outside of the dimensional specification in the labelling.

This deviation resulted from a manufacturing fixture that allowed the liner component to shift slightly out of alignment prior to backside machining.

To date, there have been no complaints or adverse events identified, in Australia or elsewhere, associated with this problem.
Instructions for Customers/Users Implant Product Alert:
1. Normal patient follow-up, per each surgeon’s standard, is recommended.
2. Do not implant any impacted product.

Recall
1. The impacted Exactech Equinoxe® Reverse Shoulder Humeral Liners have been placed on a Quality Hold and will temporarily be unavailable.
2. Exactech is liaising with the manufacturer to re-establish supply of compliant Product.
3. Exactech will remove all impacted stock from consignment locations.

This action has been finalised on 8/12/2025.
Action Commencement Date 21/08/2025
Responsible Entity Exactech Australia Pty Ltd
Contact Information 02 8882 9688 - Exactech Australia