TGA Action ID RC-2025-RN-00727-1
Type of Product Medical Device
Product Name/Description GETINGE 88-5 Washer Disinfector (all units with electrical heating)

Multiple Lot/Serial numbers
ARTG Number 409106
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Getinge has identified a problem concerning the 88-5 washer-disinfector device, whereby overheating of the heater element occurs. This problem is caused by absent or incorrect calibration of the circulation pressure sensor. This can result in the melting of insulation, cables, and hoses, and in some cases, may lead to smoke emission or fire.

This problem may cause injury, such as burns or smoke inhalation. The smoke or fire could also cause machine down time.
Instructions for Customers/Users Customers/operators are to:
- Check the serial number of Getinge 88-5 Washer disinfector.
- Verify and document the circulation pressure value or if needed by a Getinge Technician (further instructions in the customer letter).
- If the value deviates from 0.0 kPa out of allowed tolerances (±1 kPa), stop using the device immediately and contact a Getinge representative to arrange recalibration of the pressure sensor.

This action has been closed out on 28/04/2026.