TGA Action ID RC-2025-RN-00722-1
Type of Product Medical Device
Product Name/Description RAINBOW DCIP PED REUSABLE SENSOR & M-LNCS PED ADHESIVE SENSOR

Lot Numbers: 23FMK, 23EQY

Catalogue Numbers: 11171-000050, 11171-000040

GTIN: 00843997004862, 00843997002523
ARTG Number 201980, 201981
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Stryker has identified two separate labelling discrepancies involving SpO2 sensors. These discrepancies are limited to the external shipping packaging and do not affect the individual product packaging or the performance of the products.

In one instance, Rainbow DCI Adult Reusable Sensors were incorrectly labelled by Stryker as Rainbow DCIP Pediatric Reusable Sensors. As a result, customers who ordered Rainbow DCIP Paediatric Reusable Sensors (catalog number/REF 11171-000050) may have received Rainbow DCI Adult Reusable Sensors from the affected lot.

In another case, the M-LNCS Paediatric Adhesive Sensor (catalog number/REF 11171-000040) was incorrectly labelled upon receipt as catalog number/REF 11140-000080. As a result, customers who ordered a power extension cable may have inadvertently received an M-LNCS Pediatric Adhesive Sensor from the affected lot numbers.
Instructions for Customers/Users Customers are to:
· Immediately review the Customer Notification Letter and quarantine any unopened affected product in their possession.
· Inform Stryker immediately if any affected product has been loaned, sold, or otherwise disposed of.
· Expect coordination from a Stryker representative regarding the return or replacement of affected products.
· Return products within scope to the manufacturer, as part of the proposed recall action.
Action Commencement Date 20/08/2025
Responsible Entity Stryker Australia Pty Ltd
Contact Information postmarketssp@stryker.com