| TGA Action ID |
RC-2025-RN-00720-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Dolphin Imaging Software
Version: 12.0
Product Code: 148-0205 |
| ARTG Number |
287656 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class III |
| Reason for the Action |
When users select to export an image out of Dolphin Imaging and choose DICOM format, rather than the default JPG format, the software exports DICOM date of birth and acquisition date fields in the format MM-DD-YYYY, which does not comply with the required YYYYMMDD standard.
The risk of this problem occurring is very unlikely since the sequence of actions required to trigger the problem does not reflect typical workflows. Should it occur, the formatting problem does not affect device safety or essential performance and does not pose a risk to patient safety. |
| Instructions for Customers/Users |
Dolphin Imaging version 13.0 will correct this problem. The release is planned to be completed by September 2025.
In the interim customers are to: - View images and patient data directly within Dolphin Imaging rather than relying on the dates presented on the DICOM images following the export. The images and data stored in Dolphin Imaging display the dates accurately.
- Users can avoid the formatting problem entirely by using the standard viewing functionality.
Update 14/11/2025: The planned release for the updated software has been delayed and is expected to be completed by end of February 2026.
Update 04/06/2026: A software problem has been identified related to version 13.0, which was intended to correct this problem. The manufacturer has temporarily paused further distribution of version 13.0 while a corrective update is being finalised. A corrected version is in development and will be communicated once available. |
| Action Commencement Date |
15/08/2025 |
| Responsible Entity |
|
| Contact Information |
AUNZvigilance@ul.com - Emergo Australia |