| TGA Action ID |
RC-2025-RN-00714-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
ARTIS pheno
Software version VE30B/VE40B
CIB Board with firmware Rev.01 |
| ARTG Number |
273952 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Initiating a shutdown of the system can sporadically lead to a state in which the robotic C-arm stand cannot be moved after the next system startup. Other movements (e.g. table movements) are still available with reduced speed.
The system may show the following generic messages “Reduced stand/table speed” and “Limited stand movement, wait”.
In case the error occurs the C-arm stand will not move anymore. Therefore, the C-arm cannot be moved into the required position. This may result in a situation where it is necessary to cancel the clinical procedure or to begin or continue the procedure on an alternative system. |
| Instructions for Customers/Users |
Customers are to: - Test system movements after each system startup, according to the daily checks described in the operator manual. - In case the robotic C-arm stand cannot be moved, and the described system messages are displayed please call the service. - Siemens service organisation will get in contact to arrange an appointment to perform the corrective action. Completion of correction is expected by end of 2025.
This action has been closed out on 06/01/2026 |
| Action Commencement Date |
15/08/2025 |
| Responsible Entity |
|
| Contact Information |
1800 310 300 - Siemens Customer Care Centre |