TGA Action ID RC-2025-RN-00714-1
Type of Product Medical Device
Product Name/Description ARTIS pheno

Software version VE30B/VE40B

CIB Board with firmware Rev.01
ARTG Number 273952
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Initiating a shutdown of the system can sporadically lead to a state in which the robotic C-arm stand cannot be moved after the next system startup. Other movements (e.g. table movements) are still available with reduced speed.

The system may show the following generic messages “Reduced stand/table speed” and “Limited stand movement, wait”.

In case the error occurs the C-arm stand will not move anymore. Therefore, the C-arm cannot be moved into the required position. This may result in a situation where it is necessary to cancel the clinical procedure or to begin or continue the procedure on an alternative system.
Instructions for Customers/Users Customers are to:
- Test system movements after each system startup, according to the daily checks described in the operator manual.
- In case the robotic C-arm stand cannot be moved, and the described system messages are displayed please call the service.
- Siemens service organisation will get in contact to arrange an appointment to perform the corrective action. Completion of correction is expected by end of 2025.

This action has been closed out on 06/01/2026
Action Commencement Date 15/08/2025
Responsible Entity Siemens Healthcare Pty Ltd
Contact Information 1800 310 300 - Siemens Customer Care Centre