| TGA Action ID |
RC-2025-RN-00712-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Allura Xper systems
Systems: ALLURA XPER FD20 BIPLANE, Allura Xper FD10, Allura Xper FD20, Allura Xper FD20/15
Material Numbers: 722013, 722026, 722028, 722058 |
| ARTG Number |
225815 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Philips has identified that the air baffle component may be potentially missing in some Nehalem FlexVision-2 PCs within the Allura Xper series.
The air baffle directs part of the airflow inside the PC. When it is not installed, the temperature of the frame grabber card of the FlexVision-2 PC may be outside the specified temperature range, which may lead to an increased probability of failure of the frame grabber card. If this occurs, no image will be shown in the FlexVision monitor of the examination room.
From 1st of January 2022 to 12th of June 2025, Philips has received 35 complaints potentially related to this problem, but none reporting harm related to this problem.
Potential health consequences associated with the unavailability of display of X-ray imaging on the FlexVision monitor are related to a potential delay of therapy. |
| Instructions for Customers/Users |
Philips will be inspecting affected Nehalem Flexvision-2 PCs and installing the air baffle if it is confirmed to be missing.
A local Philips representative will contact customers to schedule a visit to perform this activity in affected systems.
Philips expects to start the implementation of this action in affected devices by Q4 2025.
This action has been closed out on 25/02/2026 |
| Action Commencement Date |
19/08/2025 |
| Responsible Entity |
|
| Contact Information |
1800 251 400 - Service Delivery Team |