| TGA Action ID |
RC-2025-RN-00705-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Alcon Standalone Vitrectomy Consumables
ASSY,SHIP,CONSTELLATION VFC Catalogue number: 8065750957 Lot number: 1790T0
25GA HP ENTRY SYSTEM, 6 MM 1 COUNT Catalogue number: 8065000358 Lot number: 17DR9L
27GA HP ENTRY SYSTEM, 6 MM 1 COUNT Catalogue number: 8065000359 Lot number: 17DWCY |
| ARTG Number |
141454, 148011 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
There is the potential that the Tyvek cover of some pouches within specific lots of Alcon Standalone Vitrectomy Consumables have an incomplete seal. There is a risk that the sterile barrier may be compromised.
The use of non-sterile surgical products may increase the risk of post-operative infection, which may require additional medical and/or surgical intervention.
To date Alcon has not received any reports of customer complaints or adverse events related to this problem. |
| Instructions for Customers/Users |
Customers are to: - Identify and quarantine affected stock. - Return affected stock to Alcon following the instructions in the customer letter. - Contact Alcon customer service for a credit.
This action has been closed out on 02/03/2026. |
| Action Commencement Date |
13/08/2025 |
| Responsible Entity |
|
| Contact Information |
1800 224 153 - Alcon Surgical Customer Services |