TGA Action ID RC-2025-RN-00703-1
Type of Product Medical Device
Product Name/Description FLEX Monoclonal Mouse Anti-Human Ready-to-use antibody CD20cy Clone L26 on Dako Omnis. An in vitro diagnostic medical device (IVD).

FLEX Mab a Hu CD20cy, cl L26, RTU(Omnis)
Part Number: GA60461-2
Lot Numbers: 41810164, 41700704

Mono MxH CD20cy, cl L26 RTU_Omnis_CN
Part Number: GA60461-2CN
Lot Numbers: 41781549, 41762487, 41651581

FLX Mab a HuCD20cy cl L26, RTU Omnis Jp
Part Number: GA60461-2J
Lot Numbers: 41810156, 41741883
ARTG Number 183436
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Weak staining has been observed in a few cases where GA604 was used to stain CLL/SLL tissue samples for CD20. However, this weak staining has not been observed in the majority of cases where GA604 was used to test CLL/SLL tissue.

As a reminder, the diagnosis of CLL/SLL is made in the context of the histologic appearance by Hematoxylin and Eosin (H&E) staining, supported by a panel of immunohistochemical markers (routinely including other B cell markers such as PAX 5 or CD79), ancillary studies (e.g. flow cytometry on peripheral blood or bone marrow), and relevant clinical information.

The potential for a false negative CD20 IHC result using GA604 may pose a remote risk of false negative CD20 IHC results in low CD20-expressing CLL/SLL tissue, potentially affecting diagnosis or treatment decisions.
Instructions for Customers/Users Customers are to:
1. Check their inventory for any affected products as these may exhibit weak staining that could result in false negative CD20 identification when used on CLL/SLL tissue samples.
2. Contact Agilent Technical Support for troubleshooting if weak staining is observed.

This action has been closed out on 25/11/2025.
Action Commencement Date 13/08/2025
Responsible Entity Agilent Technologies Australia Pty Ltd