Print version of this report
TGA Action ID RC-2025-RN-00701-1
Type of Product Medical Device
Product Name/Description G7 Android App version 2.3.1 (SW12299) and 2.4.0 (SW12300)
ARTG Number 169241
Action Type Product Correction
Action Level Consumer
Hazard Classification Class II
Reason for the Action The G7 Android App version 2.3.1 and 2.4.0 may stop updating estimated glucose values (EGVs) and no longer alert users. A fix for this problem was included in the G7 Android App SW12299 version 2.5.0 which was released on 9 October 2024.
This market action was undertaken prior to TGA notification.
Instructions for Customers/Users A fix for this problem was included in the G7 Android App SW12299 version 2.5.0, which was released on 9 October 2024.
Dexcom confirms that as of 14 July 2025 no users remained on the affected app version in Australia.
The problem has been effectively resolved through the release of the updated software version.

This action has been finalised on 3/11/2026.
Action Commencement Date 12/08/2025
Responsible Entity AA-Med Pty Ltd
Contact Information 1300 851 056 - Dexcom Technical Support