| TGA Action ID |
RC-2025-RN-00690-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
The PhD Hemostasis Valve
Product Codes: MAP800/B, MAP802/B
Identifiers: I3206933, I3129725, I3129726, I3200996, I3209935, I3215925
Expires: 21/04/2028, 6/02/2028, 15/04/2028, 29/04/2028 |
| ARTG Number |
169140 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Merit Medical Australia is conducting a recall of the PhD Hemostasis Valve due to a manufacturing defect. When used, the valve cap may remain depressed, leading to potential leakage. Use of this product may result in procedure inconvenience and/or haemorrhage.
To date Merit has not received any reports of patient harm or injury relating to this problem; however, Merit has received 113 complaints relating to this problem. |
| Instructions for Customers/Users |
Customers should immediately determine if any of the devices affected are within their facility, quarantine them, and discontinue use and distribution.
Customers are to immediately return all affected lots to Merit Medical Australia, per the instructions in the customer acknowledgement form for a credit on their account.
This action has been closed out on 26/11/2025 |
| Action Commencement Date |
5/08/2025 |
| Responsible Entity |
|
| Contact Information |
1300 696 374 - Merit Customer Service |