| TGA Action ID |
RC-2025-RN-00684-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Phenocell B 0.8%, Phenocell B 3% & Abtectcell III 0.8% RhD Negative. An in vitro diagnostic medical device (IVD).
Phenocell B 0.8% Product number: 26530201, Lot number: 2653271, Expiry: 25/08/2025
Phenocell B 3% Product number: 16640201, Lot number: 1664271, Expiry: 25/08/2025
Abtectcell III 0.8% RhD Negative Product number: 22230201, Lot number: 2223230, Expiry: 25/08/2025 |
| ARTG Number |
302034 |
| Action Type |
Product Alert |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Immulab has received market reports indicating that Cell 7 (Donor 2056017), which is heterozygous for E and used in Phenocell B 0.8%, has shown reduced or no reactivity with certain patient samples containing a known anti-E. The same donor was also used in the manufacture of Abtectcell III 0.8% RhD Negative and Phenocell B 3%, which are therefore also included in this urgent product alert. There is a potential patient safety risk due to weak or false-negative reactions, which may result in undetected clinically significant antibodies. This could compromise donor-recipient compatibility and increase the risk of incompatible blood transfusions in clinical settings. |
| Instructions for Customers/Users |
Customers are to be aware that: For Abtectcell III 0.8% RhD Negative users: • Exclude cell 2 (donor 2056017) from use. • Use additional suitable cells to assist with detection where required, or proceed to an antibody investigation panel. • Customers should consult internal procedures for antibody detection, identification and crossmatch and introduce any mitigation steps as appropriate.
For Phenocell B 0.8% and Phenocell B 3% users: • Exclude cell 7 (donor 2056017) from use. • Use additional suitable cells to confirm identification where required, or proceed to an alternative antibody investigation panel. • Customers should consult internal procedures for antibody detection, identification and crossmatch and introduce any mitigation steps as appropriate.
This action has been closed out on 27/11/2025. |
| Action Commencement Date |
6/08/2025 |
| Responsible Entity |
|
| Contact Information |
1800 252 215 - Immulab Enquiries |