TGA Action ID RC-2025-RN-00684-1
Type of Product Medical Device
Product Name/Description Phenocell B 0.8%, Phenocell B 3% & Abtectcell III 0.8% RhD Negative. An in vitro diagnostic medical device (IVD).

Phenocell B 0.8%
Product number: 26530201, Lot number: 2653271, Expiry: 25/08/2025

Phenocell B 3%
Product number: 16640201, Lot number: 1664271, Expiry: 25/08/2025

Abtectcell III 0.8% RhD Negative
Product number: 22230201, Lot number: 2223230, Expiry: 25/08/2025
ARTG Number 302034
Action Type Product Alert
Action Level Hospital
Hazard Classification Class II
Reason for the Action Immulab has received market reports indicating that Cell 7 (Donor 2056017), which is heterozygous for E and used in Phenocell B 0.8%, has shown reduced or no reactivity with certain patient samples containing a known anti-E. The same donor was also used in the manufacture of Abtectcell III 0.8% RhD Negative and Phenocell B 3%, which are therefore also included in this urgent product alert.
There is a potential patient safety risk due to weak or false-negative reactions, which may result in undetected clinically significant antibodies. This could compromise donor-recipient compatibility and increase the risk of incompatible blood transfusions in clinical settings.
Instructions for Customers/Users Customers are to be aware that:
For Abtectcell III 0.8% RhD Negative users:
• Exclude cell 2 (donor 2056017) from use.
• Use additional suitable cells to assist with detection where required, or proceed to an antibody investigation panel.
• Customers should consult internal procedures for antibody detection, identification and crossmatch and introduce any mitigation steps as appropriate.

For Phenocell B 0.8% and Phenocell B 3% users:
• Exclude cell 7 (donor 2056017) from use.
• Use additional suitable cells to confirm identification where required, or proceed to an alternative antibody investigation panel.
• Customers should consult internal procedures for antibody detection, identification and crossmatch and introduce any mitigation steps as appropriate.

This action has been closed out on 27/11/2025.
Action Commencement Date 6/08/2025
Responsible Entity Immulab Pty Ltd
Contact Information 1800 252 215 - Immulab Enquiries