| TGA Action ID |
RC-2025-RN-00680-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
VITROS Chemistry Products LAC Slides (300-Slide and 90-Slide Format). An in vitro diagnostic medical device (IVD).
Catalog numbers; 843 3880, 815 0112
Affected coatings: 0130 and above |
| ARTG Number |
232248 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
There is an increased incidence of calibration failures when attempting to calibrate VITROS LAC Slides on VITROS XT 3400 Chemistry Systems and VITROS XT 7600 Integrated Systems.
Investigation determined that lots from Coating 0130 and above, of VITROS LAC Slides may be susceptible to an increased occurrence of Condition Code TH4-63J during calibration, on VITROS XT 3400 and XT 7600 Systems. |
| Instructions for Customers/Users |
A parameter adjustment will be implemented beginning in ADD DRV 6339 (and above). ADD DRV 6339 will be available via e-Connectivity the week of August 4, 2025, and available via physical media the week of August 11, 2025.
Affected Customers should load ADD DRV 6339 (or above) upon receipt on their VITROS XT 3400 and/or VITROS XT 7600 System(s).
QuidelOrtho does not recommend reviewing past results, as the problem does not affect patient or quality control outcomes. Customers may consult their Laboratory Medical Director to address any concerns.
This action has been finalised on 19/10/2026. |
| Action Commencement Date |
1/08/2025 |
| Responsible Entity |
|
| Contact Information |
1800 032 359 - Technical Services Centre |