Print version of this report
TGA Action ID RC-2025-RN-00679-1
Type of Product Medical Device
Product Name/Description Lyphochek Assayed and Unassayed Chemistry Control. An in vitro diagnostic medical device (IVD)

Product codes: C-310-5, C-315-5, 313X, 731, 732, 730X

Lot numbers: 72941, 72942, 72940, 89781, 89782, 89780, 89761, 89762, 89760, 89771, 89772,?89770?

Expiry dates: 30-04-2027, 31-07-2027, 31-05-2028, 29-02-2028
ARTG Number 198600
Action Type Product Correction
Action Level Hospital
Hazard Classification Class III
Reason for the Action In specific lots, two changes have been made to the package insert:

1. The Reconstituted and Refrigerated storage stability for Bilirubin (Direct) has been revised and changed from 7 to 4 days in Lyphochek Assayed Chemistry Control.

2. Target levels for both Bilirubin and Bilirubin (Total) have been switched.
Instructions for Customers/Users Customers may continue to use the control as intended.

Obtain the revised package inserts from http://myeinserts.qcnet.com or see changes in the customer letter.

Discard all previous versions.

This action has been closed out on 26/11/2025.
Action Commencement Date 5/08/2025
Responsible Entity Bio-Rad Laboratories Pty Ltd
Contact Information 1800 224 354 - Bio-Rad Technical Support