| TGA Action ID |
RC-2025-RN-00678-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Trilogy Evo, Trilogy Evo O2, Trilogy Evo Universal and Trilogy EV300
Model numbers: IN2110X15B, IN2100X15B, IN2200X15B, AU2110X15B
F&P circuit part numbers: RT202, 900MR810 Philips part numbers: 1141426, 1141549 |
| ARTG Number |
332200 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Pressure delivery may incrementally exceed the specified limits (± 2 cmH2O + 4%) for pressure control accuracy when using the following circuit configurations with preset circuit settings configured on the device User Interface: • F&P Adult, RT202 single-limb inspiratory heated, 20 mm ID circuit, configured with the “Adult (20-22mm)” circuit setting • F&P Adult, 900MR810 reusable single-limb Evatherm circuit, configured with the “Adult/Pediatric(19mm)” circuit setting
All other critical parameters, including Tidal Volume measurements and alarms, continue to perform as expected and are not impacted by the circuit setting.
There have been no reports of this problem being observed in clinical practice or reports of it causing harm at this time. |
| Instructions for Customers/Users |
To ensure variance in pressure does not occur, perform the Circuit Calibration when using F&P adult RT202 and 900MR810.
To prevent the problem occurring customers should read the directions given in the customer letter and in the Calibration section for “Device Setup” and Clinical Assessment section for “Device Operation” of the Instructions for Use.
Philips Respironics is currently working on a solution to resolve this problem. An updated notification will be provided when a solution is available.
This action has been closed out on 21/07/2026 |
| Action Commencement Date |
11/08/2025 |
| Responsible Entity |
|
| Contact Information |
1800 251 400 - Philips Electronics Australia |