TGA Action ID RC-2025-RN-00667-1
Type of Product Medical Device
Product Name/Description PrisMax V2 and PrisMax V3 Control Unit and TherMax Blood Warmer Unit

Product Codes: 955558, 955725, 955515

All serial numbers
ARTG Number 311561, 311562
Unique Device Identifier (UDI) 07332414124359, 07332414126025, 07332414123925
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Vantive has identified three problems affecting the PrisMax Systems and TherMax Blood Warmer Units:
1. PrisMax System alarm T2309: Air Detected in Prime may be raised after priming with a low liquid level in the deaeration chamber.
2. PrisMax System alarm T0830: Blood Leak Detected may be raised when no blood leak is present, or the operator may have difficulty normalising the Blood Leak Detector (BLD) leading to PrisMax System alarms T1313: BLD Normalise Failed, T0853: Normalisation Failed, or T1205: BLD Self-Test Failure.
3. Thermax System may be unable to detect the presence of a bag on the Thermax Blood Warmer Unit leading to difficulty setting up therapy, or PrisMax System alarm T2284: Thermax Disposable Not Inserted.
The occurrence of these three problems may result in delays or interruptions to therapy, and in certain circumstances, could lead to blood loss.
Instructions for Customers/Users Customers are to:
· Continue using the PrisMax Systems and TherMax Blood Warmer Units in accordance with the Operator’s Manuals and on-screen instructions, as safe operation is maintained.
· Expect contact from a local Vantive representative once corrections become available, to determine the appropriate correction plan and schedule the necessary updates for impacted devices.
Action Commencement Date 7/08/2025
Responsible Entity Vantive Pty Ltd
Contact Information 1800 920 133 - Vantive Customer Service