TGA Action ID RC-2025-RN-00659-1
Type of Product Medical Device
Product Name/Description Evis Exera III Gastrointestinal Videoscope and Evis Exera III Colonovideoscope

Model numbers: GIF-HQ190, CF-HQ190L, CF-HQ190I

Product codes: N3802940, N3802140, N3802240

Lot numbers: 20209903, 20209905, 20210594

Multiple serial numbers
ARTG Number 355250, 355634
Action Type Recall
Action Level Hospital
Hazard Classification Class III
Reason for the Action Olympus identified holes in the sterile packaging for some GIF-HQ190 and CF-HQ190L/I endoscopes during post-sterilisation inspection. They are inappropriately labelled as “STERILE” due to the potential for holes in the sterile packaging. The holes in the sterile barrier are caused by the endoscope piercing the pouch.

The endoscopes affected do not require sterilization for safe patient use. Reprocessing standards require only high-level disinfection. The subject endoscopes were appropriately cleaned, high-level disinfected, and passed microbiological testing.

No complaints have been identified relating to this matter. Olympus is recalling unopened units out of an abundance of caution.
Instructions for Customers/Users Customers are to:
- Quarantine any affected devices that are still in the sterile packaging
- Return any of the Olympus On Demand loaner endoscopes(s) listed in Appendix A that are still in the sterile packaging

No action is required for endoscopes that have already been removed from the sterile packaging.

This action has been finalised on 18/12/2025.
Action Commencement Date 29/07/2025
Responsible Entity Olympus Australia Pty Ltd
Contact Information 1300 132 992 - Olympus Customer Operations