TGA Action ID RC-2025-RN-00656-1
Type of Product Medical Device
Product Name/Description Pinnacle Radiation Therapy Planning System

Reference: 70218

Software Version: 16.0.2

UDI: (01)00884838085602(10)16.0.2.60004

ARTG 118156 (cancelled as of 04/07/2025)
ARTG Number
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action There is a software problem affecting Advanced Region of Interest (ROI) Expansion and Contraction that results in a potential shift of the intended ROI. The software assumes the patient is always in the Head First Supine (HFS) position when transposing the Advanced Expansion to alternate positions. There is potential for overdose to nearby structure or potential underdose in the treatment plan due to a direction error on Head First Prone (HFP), Feet First Supine (FFS), Feet First Prone (FFP) Orientation Image sets.

There have been no reported adverse events related to this problem.
Instructions for Customers/Users Customers may continue to use the system in accordance with the intended use and by following the recommendations below:

- When using the HFP, FFS, FFP position while performing the Advanced Region of Interest (ROI) Expansion and Contraction, always verify the direction and extent of any ROI expansion or contraction displayed in the viewport before finalizing the treatment plan
- Please refer to the Instructions for Use (IFU), which emphasize reviewing contours created within the Pinnacle Radiation Therapy Planning System.

Update 1/10/25: A Philips representative will arrange for a label to be affixed to the workstation, instructing use of only the HFS position for the affected feature.

This action has been closed out on 14/05/2026
Action Commencement Date 28/07/2025
Responsible Entity Philips Electronics Australia Ltd
Contact Information 1800 251 400 - Philips Service Delivery Team