TGA Action ID RC-2025-RN-00654-1
Type of Product Medical Device
Product Name/Description CADD Medication Cassette Reservoirs

Product Numbers: 21-7300-24, 21-7301-24, 21-7302-24

Lot Numbers: 6037757, 6022021, 6022023, 6022026, 6037777, 6054007, 6054008, 6062701
ARTG Number 400443
Unique Device Identifier (UDI) 21-7300-24, 21-7301-24, 21-7302-24
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Certain CADD 50mL and 100mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.

Leakage from a weakened medication bag weld could potentially result in delay or interruption of therapy, under delivery of medication, exposure to infectious or toxic agents, or air embolus.

No serious injuries or deaths have been reported to ICU Medical because of this problem.
Instructions for Customers/Users Customers are to:

- quarantine affected stock to prevent further use.
- destroy/discard quarantined product following their institution's processes.
- return the completed response form to confirm disposal.

ICU Medical will provide replacements upon receipt of the completed response form.
Action Commencement Date 6/08/2025
Responsible Entity Smiths Medical Australasia Pty Ltd
Contact Information 1800 654 949 - ICU Medical Australia