| TGA Action ID |
RC-2025-RN-00654-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
CADD Medication Cassette Reservoirs
Product Numbers: 21-7300-24, 21-7301-24, 21-7302-24
Lot Numbers: 6037757, 6022021, 6022023, 6022026, 6037777, 6054007, 6054008, 6062701 |
| ARTG Number |
400443 |
| Unique Device Identifier (UDI) |
21-7300-24, 21-7301-24, 21-7302-24 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Certain CADD 50mL and 100mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.
Leakage from a weakened medication bag weld could potentially result in delay or interruption of therapy, under delivery of medication, exposure to infectious or toxic agents, or air embolus.
No serious injuries or deaths have been reported to ICU Medical because of this problem. |
| Instructions for Customers/Users |
Customers are to:
- quarantine affected stock to prevent further use. - destroy/discard quarantined product following their institution's processes. - return the completed response form to confirm disposal.
ICU Medical will provide replacements upon receipt of the completed response form. |
| Action Commencement Date |
6/08/2025 |
| Responsible Entity |
|
| Contact Information |
1800 654 949 - ICU Medical Australia |