| TGA Action ID |
RC-2025-RN-00650-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Remote-controlled R/F system FLEXAVISION F3/F4
Serial numbers: 4109C6227001, 41B4AC436001, 41B4AC438001, MP7A1444B001, MP7A14459001, MP7A14466001, MP7A14466002, MPEA63C96002, MQC353CC7005 |
| ARTG Number |
141010 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
When switching to the dual tube option's console (GSC option) for the FLEXAVISION (F3/F4 version) with the dual tube option immediately after selecting serial radiography, the X-rays are exposed with values different from the radiography conditions displayed.
The X-ray may be taken under suboptimal conditions, resulting in an image with an insufficient or excessive X-ray dose. If an image that is not suitable for diagnosis is obtained, another radiography image may be necessary. |
| Instructions for Customers/Users |
Shimadzu will update the software of affected devices when available.
In the interim, to prevent this problem when switching to the dual-tube option’s console (general radiography tube), switch off serial radiography first before selecting General radiography and confirm the radiography display is correct.
This action has been closed out on 28/10/2025 |
| Action Commencement Date |
25/07/2025 |
| Responsible Entity |
|
| Contact Information |
02 9898 2444 - Shimadzu Oceania |