TGA Action ID RC-2025-RN-00650-1
Type of Product Medical Device
Product Name/Description Remote-controlled R/F system FLEXAVISION F3/F4

Serial numbers: 4109C6227001, 41B4AC436001, 41B4AC438001, MP7A1444B001, MP7A14459001, MP7A14466001, MP7A14466002, MPEA63C96002, MQC353CC7005
ARTG Number 141010
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action When switching to the dual tube option's console (GSC option) for the FLEXAVISION (F3/F4 version) with the dual tube option immediately after selecting serial radiography, the X-rays are exposed with values different from the radiography conditions displayed.

The X-ray may be taken under suboptimal conditions, resulting in an image with an insufficient or excessive X-ray dose. If an image that is not suitable for diagnosis is obtained, another radiography image may be necessary.
Instructions for Customers/Users Shimadzu will update the software of affected devices when available.

In the interim, to prevent this problem when switching to the dual-tube option’s console (general radiography tube), switch off serial radiography first before selecting General radiography and confirm the radiography display is correct.

This action has been closed out on 28/10/2025
Action Commencement Date 25/07/2025
Responsible Entity Shimadzu Oceania Pty Ltd
Contact Information 02 9898 2444 - Shimadzu Oceania