TGA Action ID RC-2025-RN-00642-1
Type of Product Medical Device
Product Name/Description ESG-410 Electrosurgical Generator

Model number: WA91307W

Multiple serial numbers
ARTG Number 371561
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Following customer complaints, Olympus have identified the potential for the device to display the error message "E0662 System Error", which is then followed by either automatic single reboots of the device, or continuous reboot ‘loops’. Users reported that these reboots occurred both during and prior to clinical use.

Olympus has identified that this problem is caused by a previous change to an electronic component.

If a device unexpectedly reboots, this may cause a delay in starting surgical treatment or prolonged surgery due to the need to troubleshoot and/or replace the generator. There is the potential risk of bleeding which may require the clinician to use alternative surgical equipment (i.e., clamps, sutures, etc.) to maintain haemostasis.

43 complaints have been received related to this problem, with no adverse events reported.
Instructions for Customers/Users Olympus will contact customers to arrange a mutually convenient time to correct the affected devices.

Customers may continue to use their devices whilst they are waiting for Olympus to contact them to arrange the rework. Customers should continue to follow the IFU, including preparing a spare electrosurgical generator or an alternative procedure, in case of an interruption to a procedure due to an unexpected generator failure.
Action Commencement Date 22/07/2025
Responsible Entity Olympus Australia Pty Ltd
Contact Information 1300 132 992 - Olympus Customer Operations