| TGA Action ID |
RC-2025-RN-00627-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Consumables from MEDTRON AG
Product Name: ELS 200 ml syringe MU, MRS 222 XS tube system with exchangeable spikes and additional valve, Patient line 200cm with check valve
Model Number: 314626-100, 314119-100, 318198-100
Lot Number: H223603, H221609, H221617 |
| ARTG Number |
309336, 309337, 309338, 309342, 309343 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class I |
| Reason for the Action |
Currently the consumables from MEDTRON AG do not provide evidence that they fulfill the requirements for ethylene oxide residuals according to DIN EN ISO 10993-7:2022 (Biological evaluation of medical devices - Part 7: Ethylene oxide sterilisation residuals) for use in special patient populations, in particular children, infants, neonates and premature infants.
There might be a health risk for patients < 10 kg due to increased ethylene oxide residuals.
Thus, the use of the consumables is restricted to patients = 10 kg. |
| Instructions for Customers/Users |
- The instructions for use (IFU) of all affected products will be updated regarding the restricted patient group and will be included with the products and published on the website of MEDTRON AG within the scope of a product update in the next months.
- The products can still be used with the current instructions for use for patients = 10 kg.
- Customers must not use the products anymore for patients < 10 kg. If the products have already been used no further measures are required.
This action has been closed out on 06/11/2025. |
| Action Commencement Date |
11/07/2025 |
| Responsible Entity |
|
| Contact Information |
AUNZvigilance@ul.com - Emergo Australia |