| TGA Action ID |
RC-2025-RN-00615-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Affirm Prone Breast Biopsy System
Models: PBX-SYS-AFFIRM-3D, PBX-SYS-AFFIRM-2D
Serial numbers: APB1000127, APB1101326, APB1000114, APB1100983, APB1100869, APB1101156, APB1101018, APB1100851, APB1100668, APB1100409 |
| ARTG Number |
281546 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class III |
| Reason for the Action |
The Affirm Prone Breast Biopsy System is exhibiting a slightly elevated level of radiation to the side of the breast platform below the table. This radiation is only present when exposures are performed.
There is a very low likelihood that a patient or a care provider could receive slightly elevated radiation due to this problem. |
| Instructions for Customers/Users |
Hologic will contact customers to schedule a hardware update and, where applicable, a software update.
In the interim, the system may continue to be used. Customers should continue to follow the ALARA principles of radiation safety and anyone required to be near the patient/system during an exposure must wear lead PPE.
This action has been closed out on 06/01/2026 |
| Action Commencement Date |
9/07/2025 |
| Responsible Entity |
|
| Contact Information |
1800 264 073 - Hologic Technical Support |