TGA Action ID RC-2025-RN-00615-1
Type of Product Medical Device
Product Name/Description Affirm Prone Breast Biopsy System

Models: PBX-SYS-AFFIRM-3D, PBX-SYS-AFFIRM-2D

Serial numbers: APB1000127, APB1101326, APB1000114, APB1100983, APB1100869, APB1101156, APB1101018, APB1100851, APB1100668, APB1100409
ARTG Number 281546
Action Type Product Correction
Action Level Hospital
Hazard Classification Class III
Reason for the Action The Affirm Prone Breast Biopsy System is exhibiting a slightly elevated level of radiation to the side of the breast platform below the table. This radiation is only present when exposures are performed.

There is a very low likelihood that a patient or a care provider could receive slightly elevated radiation due to this problem.
Instructions for Customers/Users Hologic will contact customers to schedule a hardware update and, where applicable, a software update.

In the interim, the system may continue to be used. Customers should continue to follow the ALARA principles of radiation safety and anyone required to be near the patient/system during an exposure must wear lead PPE.

This action has been closed out on 06/01/2026
Action Commencement Date 9/07/2025
Responsible Entity Hologic (Australia & New Zealand) Pty Ltd
Contact Information 1800 264 073 - Hologic Technical Support