TGA Action ID RC-2025-RN-00610-1
Type of Product Medical Device
Product Name/Description X-ray High-Voltage Generator

Multiple product names, serial numbers and part numbers
ARTG Number 104778, 106595, 141009, 141010
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action A problem has been identified with the electromagnetic contactors located inside the cabinet of the X-ray high-voltage generator. Under certain conditions, these contacts may weld together, which could result in continued current flow to the power supply circuit, even when the device is powered off.

This could pose a risk of overheating, potentially affecting nearby components and leading to smoke or device failure.

However, all internal wiring, printed circuit boards, and external covers are made from flame-retardant materials. As a result, while overheating is a concern, the risk of any fire remains extremely low.

To date, no reports of this occurring, or any health hazards related to this problem have been received.
Instructions for Customers/Users A protection circuit will be added or replaced on all affected units.

In the interim customers are to turn off the power of the x-ray unit at the console according to the operation manual at the end of the day. Then turn off the breaker/isolator switch located near the x-ray generator unit.
Action Commencement Date 8/07/2025
Responsible Entity Shimadzu Oceania Pty Ltd