TGA Action ID RC-2025-RN-00605-1
Type of Product Medical Device
Product Name/Description Nellcor Bedside SpO2 Patient Monitoring System

Product codes: PM100N, 10005941

GTIN numbers: 10884521196728, 10884521163454, A8845211634501

IFU part numbers: PT00156509 Rev A, PT00156328 A, PT00156324, PT00156324

All serial numbers
ARTG Number 179890
Action Type Product Alert
Action Level Consumer
Hazard Classification Class I
Reason for the Action Medtronic are issuing an alert following multiple reports from customers stating that the device alarm was not heard or recognised, resulting in a delay to treatment, or lack of response to low oxygen saturation, respiratory failure, or arrythmia. There have also been reports of death during some circumstances.

While the investigation and analysis of the devices returned from the customer reports did not identify any anomalies or non-conformances with the products, clinicians are being reminded to ensure the Nellcor Bedside SpO2 Patient Monitors are set up appropriately for safe use in the correct patient setting mode and alarm volume.
Instructions for Customers/Users Clinicians are asked to:

- continue use of the device in accordance with the Operator's Manual
- ensure that the alarm volume is set to a level that is audible throughout the care environment
- respond to all alarms, irrespective of priority
- confirm that any devices used in the homecare environment where the caregiver is not the clinician are set to Homecare Mode. A device configured to Homecare Mode limits the settings available to the lay user.
- Ensure that all those who need to be aware of this alert are provided the information in the customer letter, including relevant patients/caregivers. As warned in the Operator's Manual, "Do not silence or decrease the volume of the audible alarm if patient safety could be compromised".

Medtronic will provide additional information, including updating the IFU, once this has been finalised.
Action Commencement Date 24/07/2025
Responsible Entity Medtronic Australasia Pty Ltd
Contact Information 02 9857 9000 - Medtronic Australasia