Print version of this report
TGA Action ID RC-2025-RN-00600-1
Type of Product Medical Device
Product Name/Description UniCel DxI 600/800 Access Immunoassay Analyzers. An in vitro diagnostic medical device (IVD).

UniCel DxI 800 Access Immunoassay Analyzer

DxI 800 Access Immunoassay Analyzer W/Dual Gantry

DxI 800 Access Immunoassay Analyzer W/Spot B

UniCel DxI 600 Access Immunoassay Analyzer

DxI 600 Access Immunoassay Analyzer W/Dual Gantry

DxI 600 Access Immunoassay Analyzer W/Spot B

Part numbers: 973100, A71457, A71456, A30260, A71461, A71460

Multiple serial numbers
ARTG Number 177999
Unique Device Identifier (UDI) Refer to attached UDI list
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Beckman Coulter has determined that UniCel DxI 600/800 instruments may experience an increase in pick and place gantry events. Specific events you may encounter in the Event Log include:
• “Pick and Place hit an unexpected vessel while attempting to place vessel”
• “X Home sensor in unexpected position”
• “Vessel missing or not detected during pick and place move: Wash Pick and Place”
• “Pick and place detected an unexpected vessel: Wash Pick and Place”
• “Vessel not detected for pickup: Wash Pick and Place”
• “Vessel missing or not detected during pick and place move: Incubator Pick and Place”
• "Pick and place detected an unexpected vessel: Incubator Pick and Place”
• “Vessel not detected for pickup: Incubator Pick and Place”
• “Vessel missing or not detected during pick and place move: Sample Pick and Place”
• “Pick and place detected an unexpected vessel: Sample Pick and Place”
• “Vessel not dd for pickup: Sample Pick and Place”

The above events will coincide with a system Not Ready mode. When this occurs, the system may cancel tests currently in progress.

If sufficient sample volume is unavailable to repeat a cancelled test, there may be a delay in reporting patient results.
This problem does not cause erroneous test results.
Instructions for Customers/Users When the customer encounters any of the events described, they are to:
• Perform a system initialisation to resume normal operation.
• If repeated initialisation attempts fail, contact Customer Technical Support.
• Repeat testing for any cancelled tests.
A Beckman Coulter representative will schedule the modification installation when it is available

This action has been finalised on 14/10/2026.
Action Commencement Date 7/07/2025
Responsible Entity Beckman Coulter Australia Pty Ltd
Contact Information 1800 060 878 - sales_aust_nz@beckman.com - Customer Administration Services