TGA Action ID RC-2025-RN-00599-1
Type of Product Medical Device
Product Name/Description VITEK 2 AST GN cards. An in vitro diagnostic medical device (IVD)

Product code: AST-XN24 TEST KIT 20 CARDS

Reference number: 424351

All unexpired lots
ARTG Number 196634
Unique Device Identifier (UDI) 03573026 630874
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Customers after implementing VITEK 2 AST GN cards with the updated Colistin formula (cs02n) are obtaining false resistant Colistin results primarily with multidrug resistant Acinetobacter baumannii and Pseudomonas aeruginosa.
Instructions for Customers/Users bioMérieux recommends as a short-term solution, whilst they investigate the root cause of the problem, that customers confirm any resistant colistin (cs02n) results for multidrug resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex isolates via an alternate method for all existing and future AST cards with cs02n.

If appropriate for customers laboratory, create a custom bioART rule to show when resistant colistin (cs02n) results are obtained for Pseudomonas aeruginosa and Acinetobacter baumannii complex isolates (refer to the VITEK 2 Software User Manual for instructions).

Based on individual facility policy, customers may wish to perform retrospective review of resistant results.

This action has been closed out on 18/02/2026
Action Commencement Date 4/07/2025
Responsible Entity Biomerieux Australia Pty Ltd