| TGA Action ID |
RC-2025-RN-00598-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
CardioMessenger Smart 4G
Product code: 438616
Multiple Serial numbers |
| ARTG Number |
352258 |
| Action Type |
Recall |
| Action Level |
Consumer |
| Hazard Classification |
Class II |
| Reason for the Action |
A subset of CardioMessenger Smart 4G devices were manufactured with the incorrect firmware and were unable to connect to the 4G mobile network to transmit medical messages from a BIOTRONIK active implant (i.e. Implantable Pulse Generator/Implantable Cardiac Monitors/Implantable Cardiac Defibrillators) to the BIOTRONIK Home Monitoring Service Center.
As of July 4, 2025, BIOTRONIK has received 45 reports of missing Home Monitoring transmissions, but no further consequences have been reported for patients involved. BIOTRONIK implants are not compromised, and all therapy and statistical functions are fully operational. |
| Instructions for Customers/Users |
Healthcare Professionals (HCPs) are to: · Immediately inspect all inventory for the affected CardioMessenger Smart 4G devices. If any affected units are identified, please quarantine them and refrain from using them. · Promptly contact any patients who have been provided with an affected CardioMessenger Smart 4G device. · Return affected stock to BIOTRONIK Australia Pty Ltd for proper inspection and replacement. · Coordinate with your BIOTRONIK sales representative or customer service team to arrange the return and replacement process. If you have supplied or transferred any potentially affected product to another facility or organization, please ensure they receive a copy of this notification.
Patients are advised that: · The faulty CardioMessenger Smart 4G will be exchanged with a fully functional device. · They follow the instructions in the box to activate the new device and place it where the previous one was. · They contact BIOTRONIK to request a prepaid return package, or use the one provided to return the affected device.
This action has been closed out on 23/12/2025 |
| Action Commencement Date |
14/07/2025 |
| Responsible Entity |
|
| Contact Information |
02 9497 3700 - Biotronik Customer Care |